POWER SCHEDULE
nu coeomis nrvo winrs
sacs uusr ec oc�RuwEo
HY iHE EL£LiPICAL
ervcirveeR or RewRo aeR
YAIN BREAKER REC RN� i0 M�IMNN
EPO CIRCUIi SEMENS IRPE6VICE
POWEA SOURCE REQOIREMENIS.
rvme A� M C125A)
PO PO PO A (100A) �P
B (30A)
�C 15A-�
�_ �__ )_� aatiow� ineic
aunFr vowEa rcEo.
LONIACI SIEMENS
PROJECi MPNPCER.
irXF1�i
G�li
y/t� T�
I LscuEHs-suPPuco
Munuuw wiRE sizc pOLY. A100 SVSTEM aow[a caa�
raR rcRuniu wcs CABINET CABINET Foa rna�c o�r�n.
is X�/o nwc roR c.c. ro coHrvEcr ro
'Pm' wewer. crci.
PUt SC1
ITEM p1V �ESCRIPTION
Mp � MAIN PANEL WITH CIRCUR BREAKERS FWSH OR SURFACE
MOUNTED.
M � MAIN 6REAKER MUST HAVE TRIPPING DEVICE 50 WHEN
ANY EPO IS PRESSEq THE MAIN BREAKER TRIPS.
MAIN BREAKER AMPS: 125
VOLTS PHASES NEUTRAL GROUND TOTAL WIRES
>Boi2»Y 3 , , 5 cNOTE ,>
A 1 BREAKER AMPS: 100 (FOR PU1)
VOLTS PHASES NEUTRAL GROUND TOTAL WIRES
480V 3 0 1 4(NOTE 1)
B 1 BREAKER AMPS: 30 (FOR SC1)
VOLTS PHASES NEUTRAL GROUND TOTAL WIRES
48oiz»� 3 , , 5 cNOTE ,>
C 1 BREAKER AMPS: 15 (FOR STEP-DOWN XFMR "XF1')
VOLTS PHASES NEUTRAL GROUND TOTAL WIRES
480 1(L1,L2) 0 1 3
xvt t �savn, aaov aaiMnav, izav caau�uoeo secoNonev
STEP—OOWN SINGLE—PHASE TRANSFORMER WITH PRIMNRY AND
SECONOARY Fl1SE PROTECTION FOR TABLE OIITLET POWER,
CONNECTE� TO AN A�JACEM FWSH WALL—MOUME� 15A,
125vqC UL 943 GFCI WITH BLANK FACE (NO CONTACT
OPENINGS OR NEAIA CONRGURATION) WITH LED IN�ICATION,
PUSH—i0—iESi AN� PUSH—iO—RESEf BUTfONS, AND A CLEAR
LEXAN HINGED COVER TO AVOID INADVERTENT MNNUAL TRIP.
1) ALL WIRES MUST BE SAME SIZE.
NOTE UNLESS OTHERWISE NOTED, ALL BREAKERS WILL BE 80% RATED
EPO VARIES NOtE 1— EPO CIRCUIi {f1
MAW CIRCUIT BREAKER EMERGENCV POWER OFF BUITON
WITH PROTECTNE COVER THAT PREVENTS ACCIDENTAL
ACiIVAtION. iHE EPO MUSi BE OF FAIL—SAFE DESIGN.
ALL EPO'S TO HAVE MECHANICAL LATCHING MECHANISM.
EPO MU51 BE RESEf BEPORE MAIN BREAKER CAN
RESUME OPERATION. CONTACTS AND WIRING
CONFIGURATION TO BE DESIGNED BV ELECTRICAL
ENGINEER OF RECORD.
THE EPOs MUST BE INSTALLED BY A QUALIFlED
ELECiRICAL CONiRACiOR ACCOR�ING i0 NAiIONAL
ELECTRICAL CODE, STATE AND LOCAL REGl1UTI0N5.
MEASURES SHOULD BE TAKEN TO DESIGN THE CIRCUIT
IN SUCH A WAY THAT IT WILL ALWAYS WORK WHEN THE
MEDICAL EOUIPMENi IS POWERED. iHE CUSiOMER IS
SOLELY RESPONSI6LE FOR THE IMPLEMENTATION OF THE
EPOs AND THEIR ASSOCIATE� CIRCURS AND MUST MAKE
THE FlNAL �EfERMINATION CONSIDERING ALL SITE
CONDItI0N5 AN� REGULA70RY FACi0R5.
ALL ITEMS LISTED IN THIS SCHEDULE SHALL BE SUPPLIED AND
INSTALLE� BY WSTOMER/COMRACTOR.
SIEMENS REMOTE SERVICE
ARTIS Q�Q.ZEN�ZEE CEILING MAXIMUM CABLE LENGTHS (TYPICAL EQUIPMENT�
POWER REQUIREMENTS
WIRING SYSTEM: 480Y D7V, 3 PHASE, 5-WIRE, 60 HZ.
MINIMUM POWER SUPPLY:
IF AN ON-SIiE TRANSFORMER IS REQUIREO i0 OBiAIN OPERAiING
VOLIAGE, IT MUST BE OF SUFFlCIENi CAPACIiY AND
CHARACTERISTICS TO MAINTAIN SUPPLY VOLTAGE AND IMPEOANCE
REQUIREMENTS (TRANSFORMER AND CONDUCTORS).
%-RAY GENERATOR (PU1) MOMENTARV 162 KVA
RnriNc (RnoiocRnanic ExPosuRE)
x-aar cENERntoR (aui) �oNc-tiME ia Kvn
anriNc �Fwoaoscoav�
SYSTEM CABWET (SCt) LONG-TIME B.5 KVA
RATING:
LWE IMPEDANCE 5 120 (m(])
POWER QUALIiY PARAMETERS
MAXIMUM LWE VOLiAGE VARIAiION }�09, OF SYSiEM VOLiAGE
PHASE IMBAL4NCE 29,
FREQUENCV VARIAiIOM } � HZ
POWER Sl1PPLV NOTES:
1. INCOMING POWER SUPPLIES FOR SIEMENS EpUIPMENT SHOULD
BE DE�ICAiED (BACK i0 SOURCE), ISOLA7ED AND INSULA7ED FROM
ANV OTHER E�IIIPMENT SUCH AS ELEVATORS, GENERATORS, HVAC
SVSTEMS, ETC.
2. SIEMENS HEALTHCARE REQUIRES THAT THE INCOMING POWER MEEfS
iHE POWER qUALIiV RE�UIREMENiS.
GROUNDING NOTES
EOUIPMENi GROUND CON�UCTOR i0 COMPLY WITH iHE
FOLLOWING:
1) SIZE GROUNDING WIRE TO SIEMENS EQUIPMENT PER
POWER SCHEDULE RE�UIREMENTS.
2) DERNED FROM THE ELECTRICAL SERVICE, TRANSFORMER
OR MAIN DISTRIBUTION PANEL FEEDING THE 9EMENS
EQl11PMENT.
3) RUN IN THE SAME CONDUIT, TROUGH OR RACEWAY AS iHE
PHASE CONDUCi0R5.
4) CONTINUOUS, WITH NO BREAKS OR USE OF CONDUIT,
CHAS515 OR EARTH AS THE SOLE GROUNDING PATH.
5) BONDED TO CHASSIS AND/OR CONDUIT IN ACCORDANCE
WITH THE NEC RE�UIREMENTS.
6) MINIMIZE CONNECTIONS OR TERMINALS TO ENSURE
coNnNuin ovER rHE uFE oF rHE iNsrn�wnoN.
7) AS A NORM, 1HERE SHOULD NOi BE ANY CURRENT
PRESENCE ON iHE GROUND CONDUCiOR, BUi IT IS
ACCEPiABLE i0 HAVE <SOOmA DURING OPERAiION OF 1HE
IMAGING EQUIPMENT.
TABLE POWER OUTLET SAFETY
NOTE LIFE-SUSTAINING E�UIPMENT MUST NOT BE CONNECTED TO
THE TABLE POWER OUTLEf (IF INSTALLED) W THE SIEMENS PATIENT
TABLE POWER WILL BE DISCONNECTED IF EPO BUTTON IS PRESSED.
7ABLE OUtLEf IS 120V, FUSED At 5A.
LIGHiING POWER
FROM WALL
SWITCH 10�-DJ
I SIEMENS SY51EM �� � VAC
CABMET "SCt' � >S
I � N
I I NEurw,�
� Roora ucHrs
ROOM LIGHTS �
I CONTROL REIAY I 24 VDC HOSPITAL
I I POWER POWER
I SUPPLV 100-2]] VAC
I I WARNING
�� LIGHT POWER
X-RAY ON I � g ���0-D] VAC
I WARNING LIGHT I
I caNtao� aEur� � ,uEuran�
I ••X—RAV ON••
L WARNING LIGHi
L19L5:
L 24 VDC REIAYS MUSi INCWOE N FREE—WHEELING DIO�E (P/N
1N40�2 OR EQl11VALENT).
2 ALL IiEMS PROVIDED BY CUSiOMER ELECiRICAL WNiRACiOR EXCEPi
WNTACTS INSIDE 9EA1EN5 "SC1" CABWEf (REIAS INSIDE DOTfED
LWES).
3. ALL WIRING THAT CONNECTS ID SIEAIENS �SC1" CABWET A1115T BE
14-18 AWG STRAN�E� WIRE.
4. 4 WIRES LA6ELED "24 VOC', "ROOA1 LIGNTS" AND "X—RAY ON" SHOlILO
BE SENT TO SIEMENS "SC1" CABMEf.
S 24 VDC RELAYS ARE TO BE SELECTED BY ELWTRICAL CONTRACTOR TO
HANOLE THE VOLTAGE AND AMPERAGE OF LIGHTING CURWIT.
6. IF NEE�E�, A SWIiCH i0 BLOCK RA�IAiION CAN BE INSiALLE� IN
SERIES WITH THE DOOR CONTnCT.
' THE SWiTCH (24 VDC / 20 Mn) IAUST BE PROVIDED ON SiTE.
' CONTROL USING +24 V FROM THE SVSTEM CONTROL CABINET
' PLAN THE SWITCH 50 THAT UNINTENTIONAL OPERNTION IS NOT
POSSIBLE.
* THE SWITCH MlIST BE PROVIDED WITH AN APPROPRIATE SVMBOL (OR
�ESIGNAiION) FROM WHICH iHE FUNC110N CAN BE RECOGNIZE�.
WARNING LIGHT/ROOM LIGHT SCHEMATIC
LIGHTING DETAIL I �o�E
CONTRACTOR SUPPLIED CABLES
Q FROM VIA TO DESCRIPTION REMARKS
PANEL 1 MP ELECTRiCAL CONTRACTOR TO SIZE PLUS GROUND SEE "POWER SCHEDULE"
MP 2 PU1 3�'2, 1�2 GROON� AN� CONNECT SEE "POWER SCHE�IILE'
MP 3 SC1 3N6, 1$6 NEUTRNL, t�y6 GROUNO ANO CONNECT SEE "POWER SCHEDULE"
MP 4 EPO pN�p, PWS GROUN� SEE "POWER SCHE�ULE"
EPO 5 EPO pk�p, PWS GROUN� EMERGENCV POWER
SC1 11 WL 2k14-18 AWG SEE "LIGHTING DEfAIC' SHEEf
E-501
SC1 12 �5 24V SIGNAL, 2�'14-19 AWG �OOR SWITCH
PANEL 13 MAVIG 120V / 1fiA POWER, 2N13, PWS GROUN� lAAVIG LAMP
XFMR
MAVIG 14 SU pN�3 (+/—), PWS GROUN� MAXIMUM LENGiH 49'
%FMR
TO ENSl1RE THE 11PTIME OF VOl1R SVSTEM Dl1RING THE WARRANN
PERIOD (AND BEVOND WITH A SERVICE AGREEMENT), 9EMEN5 REMOTE
SERVICES (SRS) REOUIRES REMOTE LOCAL AREA NEIWORK ACCE55 TO
SIEMENS SYStEMS.
THE PREFERRED CONNECTION MEfHOD IS (VPN) VIRTUAL PRNATE
NEIWORK (WHERE iHE CU40MER HAS AVAILABLE A VPN CAPA6LE
FIREWALL OR OiHER VPN APPLIANCE). 1HI5 MEfHO� PROVI�ES iHE
POSSIBILITV FOR REMOTE SVSTEM �IAGNOSTICS WITHOUT ADDITIONAL
HARDWARE. PLEASE CONTACT SIEMENS REMOTE SERVICES
(B00-888-9EM) TO �EfERMWE IF THIS MEfHOD IS SUITABLE FOR
YOUR 517E.
�F CUSTOMER �MZ ,
'IMAGING �EVICE �
ISINGLE HOST IP I
�OR IP SUBNEf LI57I
� I
��I
I O I rSIEMENS REMOtE1
I I IrSERVICES DMZ �
� I I SRS
�______J I ACCESS �
SERVER 1 �
QSCO i �
J�45 1 �
� ROUTER � �
rl19LIC � ���� I
LNTERNEfJ I
FlREWALL
CUSiOMER L — — — — — J
VPN SERVER
NETWORK REQUIREMENT
A GIGABIT NEIWORK IS REpUIRED FOR ADEpUATE IMAGE �ATA
TRANSFER SPEED BERNEEN THE IMAGER AND 3� RECONSTRUCTION
WORKStAtION. WORV(PLOW AND CLINICAL NEEDS DEMAND 3� IMAGES
BE AVAILABLE FOR REVIEW BY CLINICAL STAFF IMMEDINTELY 11PON
ACOUISRION.
AXIS IMAGE SYSTEM
(REMorE coNFicuRnrioN)
STANDARD
r► uvE oisauv . ss �vin cai�
6'-6' MONROR
� ACE (KEVBOAR� ANDffiS (VIA CRt)
MOUSE WNNECi 10 ACE)
OPTIONS:
6'-6"� REFERENCE DISPWV f65 (VIA CR1)
MONITOR
6'-6" y HAND SWITCH
6'-6"� CONTROL MODl1LE5
6'-6" y ECC
13'� EMERGENCV SiOP
5-6"� FOOT SWRCH
6'-6� pV� RECOR�ER � 65
5-11'
� ihT MONIiOR FOR
9�-�0"� pVD RECOR�ER
CONTROL ROOM
z SYSTEM CONNECTIONS 5°"`E:
NONE
POWER QUALITY
POOR POWER WILL ALiER EQUIPMENT PERFORMANCE
IT IS M THE CUSTOMER'S INTEREST THAT THE ELECTRICAL
CONiRACiOR BE RESPONSIBLE FOR iESiING AN� VERIFYING iHAi
THE EpUIPMENT POWER SUPPLV COMPLIES WITH THE SIEMENS
SPEQFlCATIONS.
0/O.UN/dE tlINC
REV. 21
RELEASED FOR CONSTRl1CTI0N
Su6jec��o compliance with allregula�ions
o4S�a�eantlLocalCodes N T I O N
City of Carmel
`�s��� �\o,v�N�n��
�
onrE: 11 /22/19
�.[ i ncFn r [-r, [ [ r [ a r r- [rr- °r i rnr - [ -.::.au. i-r ar[ r _oys F ace �[u i r� ti i i .. i r[ i ni. i rr
. �. � � r rr� ,.. . . - r ., �� i_e � ia� v F�rc -ac� i o�- i;� r r « o. i i .-_ ioi r<. ua , ur�
[ r - r r - r i v. r [ -- r ,-� r =i _ r e[c ,,r:�
' R � y�,: � �_s.- --,�_ _ cR � _R,� r - -1 -` -' - � - - - -�� � � _ , �Pc.- -
SIEMENS
IU HEALTH NORTH HOSPITAL
09 OB t9 RE�ocnre nnnceR nNo 1t700 NORTH MERIDIAN STREET, CARMEL, IN. 46032
� � CoouNG UNIT EXISTING ROOM - ARTIS Q CEILING